Table X.

Speckle Statistics For Physical Phantom 10-mm-Diameter Cysts (N = 5)

CR (dB) SNRs CNR (dB) GCNR
DNNpoint 35.3 ± 0.9 1.72 ± 0.03 4.56 ± 0.15 0.984 ± 0.002
DNNcyst 34.3 ± 0.6 1.76 ± 0.01 4.73 ± 0.07 0.969 ± 0.003
DNNTMPcyst 45.6 ± 1.6 1.78 ± 0.03 4.95 ± 0.17 0.989 ± 0.002
DNNpoint+cyst 48.8 ± 0.4 1.66 ± 0.04 4.34 ± 0.19 0.983 ± 0.003
DNNpoint+TMPcyst 44.7 ± 1.4 1.74 ± 0.02 4.74 ± 0.13 0.986 ± 0.002
DAS 19.6 ± 0.4 1.70 ± 0.03 3.58 ± 0.19 0.921 ± 0.004