Value 95% Confidence
LFD performance against PCR
Sensitivity 78% 40% to 97%
Specificity 100% 85% to 100%
Positive Predictive Value 100%
Negative Predictive Value 92% 77% to 98%
LFD performance against Clinical Diagnosis
Sensitivity 82% 48% to 98%
Specificity 97% 86% to 100%
Positive Predictive Value 90% 56% to 98%
Negative Predictive Value 95% 84% to 98%
LFD Result time (range) 10 min (1–30 min)
Median PCR/NAAT Result Time (range) 9 h (01.21 h to 22.56hrs