Table 3.

BU SARS-CoV-2 Test LoD Confirmation Study using replicates of 20 Positive Residual Patient Samples. LoD established at 10.6 copies per microliter.

Concentration Result
21.3 copies/µl 20/20
10.6 copies/µl 20/20
5.3 copies/µl 17/20