Published in final edited form as: Anal Chem. 2022 Jun 3;94(23):8302–8308. doi: 10.1021/acs.analchem.2c00748

Table 1.

Calibration-free ISEs Based on the PEDOT:PEDOT-S Back Contacta

electrode type: charged K+ ISE Ca2+ ISE
slope (mV/dec) 57.2 ± 0.2 28.5 ± 0.3
detection limit (μM) 1.68 ± 0.18 0.45 ± 0.01
logKA,Bpot −3.53 ± 0.02 vs Na+ −3.82 ± 0.01 vs Mg2+
−4.13 ± 0.21 vs Ca2+ −5.26 ± 0.12 vs K+
a

Characterization of a batch of three calibration-free ISEs over three replicate calibrations. The Ca2+ ISE (n = 9) and the K+ ISE (n = 9) were calibrated versus an Ag/AgCl reference electrode under optimized solution conditions.