Table 1.

Scan parameters per participating centre

Hospital MUMC+ Nijsmellinghe AMC Maasstad
Scanner 1.5 T Philips Intera 1.5 T Siemens Magnetom 1.5 T Siemens Magnetom 1.5 T Siemens Magnetom
Sequence Ultrafast GE T1 VIBE T1 VIBE T1 VIBE
Contrast Yes Yes Yes Yes
Scan mode 3D 3D 3D 3D
Repetition time (TR) (ms) 7.8 4.73 5.9 3.2
Echo time (TE) (ms) 3.9 2.17 2.44 1.28
Flip angle (degrees) 10 10 20 10
AVG acquisition time (TA) (min) 14:52 12:42 15:39 07:12
Bandwidth (BW) (Hz) 181.8 390 240 521
Acquisition voxel (mm) 0.95 × 0.95 × 3.00 0.91 × 0.91 × 1.80 0.8 × 0.8 × 0.8 1.0 × 1.0 × 6.00
Reconstructed voxel (mm) 0.95 × 0.95 × 1.50 0.91 × 0.91 × 1.80 0.8 × 0.8 × 2.0 0.9 × 0.9 × 1.5
Number of slices 750 (5 × 150) 768 (3 × 256) 537 (3 × 176) 864 (6 × 144)
Acquisition matrix 380 × 266 230 × 256 263 × 350 400 × 313
FoV 400 490 500 400
Fat suppression Yes Yes Yes Yes
Cardiac synchronization (ECG) Yes No No Yes
Breath hold No No No No