Table 1.
MRI sequencing parameters for DWI acquisition on Elekta Unity MR-Linac.
| Parameters | Rectal DWI sequencing |
|---|---|
| Field of view (mm) | AP = 420 |
| RL = 420 | |
| FH = 120 | |
| Recon. Voxel (mm) | |
| AP | 1.75 |
| RL | 1.75 |
| TR (ms) | 4483 |
| TE (ms) | 81 |
| DELTA/delta (ms) | 40.7/20.3 |
| Fat Suppression | STIR |
| EPI factor | 55 |
| Parallel imaging factor | 2.2 |
| Section thickness (mm) | 4 |
| Direction of motion probing gradients | Isotropic |
| b-factors (s/mm2) | 0, 30, 150, 500 |
| b-factor averages | |
| b = 0 | 8 |
| b = 30 | 8 |
| b = 150 | 8 |
| b = 500 | 16 |