Table 21.

nanoLC Parameters

Sample pickup and Loading
Sample Pickup
Volume 2 μL
Flow 10 μL/min
Sample Loading
Volume 6 μL
Flow 1 μL/min
Max. Pressure 700 Bar
Solvents A: water/acetonitrile + formic acid (98/2 + 0.1 % by volume) B: water/acetonitrile + formic acid (80/20 + 0.1 % by volume)
Gradient
Time Duration Flow (nL/min) %B
0:00 0:00 300 5
2:00 2:00 300 5
107:00 105:00 300 25
122:00 15:00 300 40
123:00 1:00 300 95
124:00 1:00 300 95
140:00 16:00 300 95
Pre-column and Analytical Column
Pre-column Equilibration
Volume 3 μL
Flow 3.00 μL/min
Max. Pressure 700 Bar
Analytical Column Equilibration
Volume 6 μL
Flow - μL/min
Max. Pressure 700 Bar
*

Custom wash used for Autosampler and column heating controlled externally on Easy-Spray Ion Source. Column is kept at 45 °C throughout analysis.