Skip to main content
. 2022 Jun 14;12(6):412. doi: 10.3390/bios12060412

Table 1.

Influence of excess of human albumin, leptin, CEA, CA-125 and metalloproteinase-2 (1000:1) on the results of determination of IL-6 concentration by the biosensor with antibody as the receptor (biosensor (a)).

Interferent IL-6 Spike
(pg mL−1)
Found IL-6
(pg mL−1)
Recovery
(%)
CEA 20.0 19.4 ± 2.68 97.0
CA 125 20.0 19.7 ± 1.31 98.5
Leptin 20.0 20.6 ± 3.05 103
MMP 2 20.0 20.5 ± 3.94 102
Albumin 20.0 19.0 ± 5.88 97.9