Test results of ring trial samples in HeV-G ELISA. When applying the cut-off PP values based on ROC analysis with the highest combined DSe and DSp of 40, two dilutions of positive serum samples were identified as questionable and would be repeated by Western Blot or IF analysis.

Sample ID Sample Information PP Result Interpretation
neg control 8.55 neg
pos control 100.00 pos
RTS 1 infected horse number 1; 1/100 47.72 pos correct
RTS 2 vaccinated horse 76.57 pos correct
RTS 3 infected horse number 1; 1/150 35.28 neg questionable
RTS 4 infected horse number 2; 1/60 43.59 pos correct
RTS 5 negative horse 8.21 neg correct
RTS 6 infected horse number 2; 1/40 42.36 pos correct
RTS 7 infected horse number 2; 1/60 31.43 neg questionable
RTS 8 infected horse number 2; 1/40 59.40 pos correct
RTS 9 negative horse 10.25 pos correct