Table 2.

Method validation parameters and their definitions as ipplied to the quantitation of soluble aggregates by SV-AUC.

Method validation parameter Definition


Specificity Ability to resolve aggregates as a population of c(s) species with s-values greater than that of the monomer


Accuracy Agreement of the measured value with the accepted true or reference value


Precision Closeness between multiple aggregate measurements


Limit of Detection (LOD) Lowest amount of aggregate which can be reliably detected


Limit of Quantitation (LOQ) Lowest amount of aggregate which can be reliably quantified


Linearity Ability of SV-AUC to detect and quantify soluble aggregates directly proportional to their concentration in the sample


Range Interval between the highest and lowest amount of soluble aggregates where the method has the appropriate degree of precision, accuracy, and linearity


Robustness Ability of the SV-AUC method to provide reliable quantitation with respect to small variations in method parameters