Loading sample on the reference side and vice versa |
Erroneous absorbance readings, potentially incorrect lamp intensity set |
Analyst error |
|
|
|
Cell leakage during the run |
Vacuum decrease/jumps, moving meniscus, fitting artifacts, significantly unequal column heights after the run |
Inappropriate cell sealing |
|
|
|
Cell leakage at the start of a run at 60k/50k rpm |
Moving meniscus, vacuum decrease, lack of expected data type during fitting |
Inappropriate cell sealing |
|
|
|
Contamination from another sample |
Unexpected species in the c(s) distribution |
Inappropriate cleaning of cell and centerpiece |
|
|
|
Unstable temperature during the run |
Best-fit meniscus is far off the optically discerned meniscus; poor fits in initial scans |
Insufficient temperature equilibration time |
|
|
|
High solute (reference) absorbance |
Low signal-to-noise, poor fits with high RMSD |
Absorbing excipients present |