Table 1.
Characteristics | BioFire ® Respiratory Panel 2.1 - plus | QIAstat-Dx Respiratory SARS-CoV-2 Panel |
---|---|---|
Pathogens detected | - Viruses: Adenovirus, Coronaviruses 229E, HKU1, NL63, and OC43, SARS-CoV-2, Human Metapneumovirus A/B, Influenza A (with differentiation of Influenza A/H1, A/H3, A/H1–2009), Influenza B, Parainfluenza virus 1,2 3 and 4, MERS-CoV, Rhinovirus/Enterovirus, Respiratory Syncytial Virus A/B, SARS-CoV-2 - Bacteria: Bordetella pertussis, Bordetella parapertussis, Chlamydia pneumoniae, Mycoplasma pneumoniae |
- Viruses: Adenovirus, Bocavirus, Coronaviruses 229E, HKU1, NL63, and OC43, SARS-CoV-2, Human Metapneumovirus A/B, Influenza A (with differentiation of Influenza A/H1, A/H3, A/H1–2009), Influenza B, Parainfluenza virus 1,2 3 and 4, Rhinovirus/Enterovirus, Respiratory Syncytial Virus A/B, SARS-CoV-2 - Bacteria: Bordetella pertussis, Mycoplasma pneumoniae, Legionella pneumophila |
Turnaround time | 45 min | 60 min |
Sample input volume | 300 µL | 300 µL |
Report of Ct value | No (melting curves) | Yes |