TABLE 1.

Performance of DPP Syphilis Screen & Confirm with visual operator inspection and microreader analysis (cutoff of 9.0) against reference comparator tests

Assay and detection methoda No. of samples with:
Agreement (95% CI) (%) Sensitivity (95% CI) (%) Specificity (95% CI) (%) Kappa coefficient (95% CI)
True-positive results True-negative results
Chembio treponemal component n = 304 n = 79
 Operator reading 261 79 88.7 (85.6–91.9) 85.9 (81.4–89.6) 100.0 (95.4–100.0) 0.71 (0.63–0.79)
 Microreader 253 79 86.6 (83.2–90.1) 83.2 (78.5–87.2) 100.0 (95.4–100.0) 0.67 (0.59–0.75)
Chembio nontreponemal component n = 242 n = 141
 Operator reading 167 125 76.2 (71.9–80.5) 69.0 (62.8–74.8) 88.7 (82.2–93.4) 0.53 (0.44–0.61)
 Microreader 159 126 74.4 (70.0–78.8) 65.7 (59.4–71.7) 89.4 (83.1–93.9) 0.50 (0.41–0.58)
a

Reference tests were the TPPA test for treponemal antibodies and the RPR test for nontreponemal antibodies.