TABLE 2.

Performance of FilmArray Pneumonia Panel plus (FilmArray PP) in nasopharyngeal (NP) swabs for detection of S. pneumoniae, H. influenzae, and M. catarrhalis in adults with community-acquired respiratory tract infections (n = 118 patients) compared to various reference standards: FilmArray PP on a good-quality lower respiratory tract (LRT) sample with a positive cutoff value of >104 copies/mL or with the standard microbiology results (lower respiratory tract sample culture, hemocultures, and S. pneumoniae urinary antigen)a

Target organism Reference standard No. of positive samples
PPA NPA OPA LR+ LR− PPV NPV
NP Reference standard
S. pneumoniae LRT PCR FilmArray PP 11 19 0.58 1 0.93 Infinite 0.42 1 0.93
Standard microbiology 11 12 0.58 0.96 0.92 15.46 0.43 0.64 0.95
H. influenzae LRT PCR FilmArray PP 27 44 0.61 1 0.86 Infinite 0.39 1 0.81
Standard microbiology 27 19 0.74 0.87 0.85 5.61 0.3 0.52 0.95
M. catarrhalis LRT PCR FilmArray PP 9 14 0.57 0.99 0.94 59.43 0.43 0.89 0.94
Standard microbiology 9 0 0.92 0.92 0 1
a

Positive percent agreement (PPA), negative percent agreement (NPA), overall percent agreement (OPA), positive likelihood ratio (LR+), negative likelihood ratio (LR−), positive predictive value (PPV), negative predictive value (NPV).