Table 1. Studies included in the systematic review and meta-analysis on the diagnostic accuracy of rapid tests for Ebola virus disease.

Study Country Sample size, no. Study design Reference standarda Specimen Year of specimen collection Industry funded Index test No. of samples
% (95% CI)
Prevalence, %
True positive False positive False negative True negative Reported sensitivity Reported specificity
Moran et al., 202015 Guinea 205 Cross-sectional RT–PCR (altona) Stored serum 2014–2015 No Fever Panel Antigen System 98 9 11 87 90 (83–95) 91 (83–96) 53
Ebola Antigen System 84 8 25 88 77 (68–84) 92 (84–96) 53
Ebola–Malaria Antigen duplex system 85 4 24 92 78 (69–85) 96 (89–99) 53
Makiala et al., 201916 Democratic Republic of the Congo 928 Cross-sectional GeneXpert® Ebola Whole blood 2018 No QuickNaviTM-Ebola 68 2 12 846 85 (75–92) 100 (99–100) 9
Wonderlyet al., 201917 Sierra Leone 428 Cross-sectional RT–PCR (altona) Whole blood 2015 No, but tests donated DEDIATEST Ebola 0 0 0 98 100 (96–100)
RT–PCR (altona) Stored plasma 2015 No, but tests donated SD Ebola Zaire Ag 109 2 20 196 85 (77–90) 99 (96–100)
RT–PCR (Trombley) Stored plasma 2015 No, but tests donated SD Ebola Zaire Ag 110 1 46 170 71 (63–78) 99 (97–100)
RT–PCR (altona) Stored plasma 2015 No, but tests donated ReEBOV™ Antigen Rapid Test kit 123 39 9 159 93 (88–97) 80 (74–86)
RT–PCR (Trombley) Stored plasma 2015 No, but tests donated ReEBOV™ Antigen Rapid Test kit 133 29 23 142 85 (79–90) 83 (77–88)
RT–PCR (altona) Stored plasma 2015 No, but tests donated One step Ebola test 125 39 2 158 98 (94–100) 80 (74–86)
RT–PCR (Trombley) Stored plasma 2015 No, but tests donated One step Ebola test 138 26 16 144 90 (84–94) 85 (78–90)
RT–PCR (altona) Stored plasma 2015 No, but tests donated DEDIATEST EBOLA 101 31 26 167 80 (72–86) 84 (79–89)
RT–PCR (Trombley) Stored plasma 2015 No, but tests donated DEDIATEST EBOLA 108 24 46 147 70 (62–77) 86 (80–91)
Colavita et al., 201818 Sierra Leone 210 Cross-sectional qRT–PCR (altona) Stored residual plasma samples 2014–2015 Yes EBOLA Ag K-SET 93 2 12 103 89 (81–94) 98 (93–100) 50
Gallais et al., 201719 Guinea 137 Cross-sectional qRT–PCR (altona; and Weidmann) Whole blood 2015 No eZYSCREEN® 32 1 17 87 65 (50–78) 99 (94–100) 36
144 Cross-sectional qRT–PCR (altona; and Weidmann) Serum 2015 No eZYSCREEN® 41 0 14 89 75 (61–85) 100 (96–100) 38
Phan et al., 2016b,20 Liberia 290 Cross-sectional rRT–PCR (Trombley) Plasma 2014–2015 No NMRC Ebola virus lateral flow immunoassay 88 (75–94) 98 (95–99)
237 Cross-sectional rRT–PCR (Trombley) Postmortem oral swabs 2014–2015 No NMRC Ebola virus lateral flow immunoassay 89 (57–98) 96 (93–98)
Boisen et al., 201621 Sierra Leone 176 Cross-sectional Ebola virus-specific qRT–PCRc Stored plasma 2014 Yes ReEBOV™ Antigen Rapid Test kit 72 2 44 58 62 (53–71) 97 (89–100) 53
Broadhurst et al., 201522 Sierra Leone 105 Cross-sectional RT–PCR (altona) Plasma (for the reference test) and capillary blood (for the rapid tests) 2015 No, but tests donated ReEBOV™ Antigen Rapid Test kit 28 6 0 71 100 (88–100) 92 (84–97) 27
277 Cross-sectional RT–PCR (altona) Whole blood 2015 No, but tests donated ReEBOV™ Antigen Rapid Test kit 45 18 0 214 100 (92–100) 92 (88–95) 16
Walker et al., 201523 Sierra Leone 131 Cross-sectional RT–PCR (altona) Capillary blood 2015 No Defence Science and Technology Laboratory Ebola virus disease rapid diagnostic test 15 9 107 0 100 (78–100) 92 (86–96) 12

CI: confidence interval; PCR: polymerase chain reaction; RT–PCR: reverse transcription PCR; rRT–PCR: real-time RT–PCR; qRT–PCR: quantitative RT–PCR.

a altona refers to RealStar Filovirus Screen RT–PCR Kit 1.0 (altona Diagnostics GmbH, Hamburg, Germany), Trombley refers to the Trombley assay,14 GeneXpert® Ebola (Cepheid, Sunnyvale, United States of America) is a specific RT-PCR that targets the glycoprotein and nucleoprotein genes, Weidmann refers to the Weidmann technique (that is, a quantitative one-step RT–PCR).19

b We were unable to accurately reconstitute the two-by-two table used to calculate specificity and sensitivity.

c RT–PCR specific for Ebola (while not specifically mentioned in the article, the authors cite the Trombley assay in their methods).