Key analytical figures of merit calculations used for evaluation of method performance. Linear dynamic range is defined as the range that a linear relationship was observed in our calibration curves, listed out as the minimum and maximum concentration of standard that was reliably analyzed.

Analyte Linear Dynamic Range (ng mL−1) r LOD (ng mL−1) LOQ (ng mL−1) Precision (%CV)
BAC-10 0.0179–35.8 0.999 0.025 0.080 4.01
BAC-12 0.0678–1360 0.997 0.009 0.030 1.32
BAC-14 2.46–369 0.986 5.53 18.4 63.1
BAC-16 0.0454–90.8 0.999 0.030 0.100 2.50
BAC-18 0.00124–24.8 1.00 0.003 0.010 2.62
AEB-12 0.0610–1220 0.998 0.016 0.050 2.47
AEB-14 0.0248–496 0.999 0.007 0.020 1.03
X 0.100–2000 0.993 2.10 7.01 16.1