Table 5. A comparison of the results measured by WAM and TOPCON before cycloplegia with the subjective refraction results after cycloplegia.
Group | SD (D) | SE (D) | J0 (D) | J45 (D) | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
95% CI | P | 95% CI | P | 95% CI | P | 95% CI | P | ||||||||
WAM vs. TOPCON | 0.011* | 0.021* | 0.172 | 0.156 | |||||||||||
WAM | −1.607±2.474 | −1.974 to −1.240 | −2.142±2.563 | −2.522 to −1.762 | 0.456±0.497 | 0.383 to 0.530 | −0.009±0.191 | −0.037 to 0.020 | |||||||
TOPCON | −1.749±2.551 | −2.127 to −1.370 | −2.272±2.668 | −2.668 to −1.876 | 0.481±0.542 | 0.400 to 0.561 | −0.027±0.198 | −0.057 to 0.002 | |||||||
WAM vs. subjective refraction | <0.001* | <0.001* | 0.834 | 0.207 | |||||||||||
WAM | −1.607±2.474 | −1.974 to −1.240 | −2.142±2.563 | −2.522 to −1.762 | 0.456±0.497 | 0.383 to 0.530 | −0.009±0.191 | −0.037 to 0.020 | |||||||
Subjective refraction | −1.297±2.608 | −1.684 to −0.910 | −1.778±2.694 | −2.177 to −1.378 | 0.460±0.518 | 0.383 to 0.537 | −0.023±0.171 | −0.049 to 0.002 | |||||||
TOPCON vs. subjective refraction | <0.001* | <0.001* | 0.098 | 0.693 | |||||||||||
TOPCON | −1.749±2.551 | −2.127 to −1.370 | −2.272±2.668 | −2.668 to −1.876 | 0.481±0.542 | 0.400 to 0.561 | −0.027±0.198 | −0.056 to 0.002 | |||||||
Subjective refraction | −1.297±2.608 | −1.684 to −0.910 | −1.778±2.694 | −2.177 to −1.378 | 0.460±0.518 | 0.383 to 0.537 | −0.023±0.171 | −0.049 to 0.002 |
*, P<0.05 was considered statistically significant. SD, spherical diopter; SE, spherical equivalence; J0, the axial 0°/180° representing with/against the rule of astigmatism; J45, the axial 45°/135° representing the oblique axis of astigmatism; , mean ± standard deviation; CI, confidence interval; D, diopter.