Table 1. Winston-Lutz QA outcomes for three beamline setups.

The distance of the ballbearing to the beam isocentre was captured via radiochromic film in the X, Y and Z axes. Measurement uncertainties (as described in Section 2.4.1) are shown for each axis; the three errors are combined in quadrature to give the total uncertainty in the 3D measurement.

  Distance to isocentre (mm)
  X (±0.050) Y (±0.050) Z (±0.050) 3D (±0.087)
Beamline setup 1 0.013 0.037 0.167 0.171
Beamline setup 2 0.041 0.032 0.049 0.074
Beamline setup 3 0.331 0.543 0.780 1.007

This result highlights an error in the beamline setup.