Skip to main content
. 2021 Sep 28;2(3):214–235. doi: 10.12336/biomatertransl.2021.03.005

Table 1. Mechanical performance of commercial pure (CP) titanium, stainless steel and cobalt–chromium alloys for surgical implant application.

Yield strength (MPa) Ultimate tensile strength (MPa) Elongation (%)
CP Ti
 Grade 1 170 240 24
 Grade 2 275 345 20
 Grade 3 380 450 18
 Grade 4 483 550 15
Stainless steel
 18Cr-14Ni-2.5Mo
  Annealed 190 490 40
  Cold-worked 690 860 12
Co-28Cr-6Mo
 Annealed 517 897 20
 Hot-worked 700 1000 12

Note: Superscript a indicates a different grade of CP Ti which often means a different oxygen content. Data are from ASTM International.16-18