Table 1. Compliance Values in Artificial Grafts on the Market and Autologous Grafts.
| graft name | polymer | wall thickness (μm) | pressure (mmHg) | compliance (%/100 mmHg) | ref |
|---|---|---|---|---|---|
| Gore-tex | ePTFE | 454 ± 4 | 50–90 | 0.0036 ± 0.001 | (55) |
| 80–120 | 0.0034 ± 0.0004 | ||||
| 110–150 | 0.0033 ± 0.0005 | ||||
| Zeus, Orangeburg | ePTFE | 100 | 80–120 | 3.50 ± 1.4 | (56) |
| SW-ePTFE woven Vascutek Co. | ePTFE | 694 ± 14 | 50–90 | 0.2 | (57) |
| 80–120 | 0.6 | ||||
| 110–150 | 0.6 | ||||
| ePTFE grafts, Impra, UK | ePTFE | 30–100 | 1.2 ± 0.3 | (58) | |
| CPU, CardioTech International, UK | PC | 30–100 | 8.1 ± 0.4 | (58) | |
| PU | |||||
| Silkothane | PCU/SF | 335 ± 68 | 80–120 | 4.8 ± 1.0 | (59) |
| Dacron | PET | 80–120 | 1.9 | (28) | |
| polyester (woven) | 80–120 | 2.3 | |||
| polyester (knitted) | |||||
| uncrimped knitted Dacron graft, Sulzer Vascutek, UK | PET | 30–100 | 1.8 ± 1.2 | (58) | |
| young coronary artery | human | 50–90 | 8.92 ± 7.67 | (60) | |
| 80–120 | 3.39 ± 6.03 | ||||
| 110–150 | 2.69 ± 0.89 | ||||
| old coronary artery | human | 50–90 | 4.01 ± 2.08 | (60) | |
| 80–120 | 4.06 ± 3.16 | ||||
| 110–150 | 3.42 ± 1.72 | ||||
| mammary artery | human | 50–90 | 21.05 ± 26.04 | (60) | |
| 80–120 | 5.22 ± 5.13 | ||||
| 110–150 | 1.79 ± 1.96 | ||||
| 80–120 | 11.5 ± 3.9 | (61) | |||
| radial artery | human | 50–90 | 5.66 ± 2.12 | (60) | |
| 80–120 | 2.05 ± 3.68 | ||||
| 110–150 | 1.06 ± 4.73 | ||||
| external iliac artery | human | 30 to 100 | 8.0 ± 5.9 | (58) | |
| saphenous vein | human | 50–90 | 2.60 ± 1.72 | (60) | |
| 3.46 ± 1.21 | 80–120 | 1.77 ± 0.20 | |||
| 110–150 | 0.67 ± 0.89 | ||||
| 30 to 100 | 5.0 ± 6.0 | (58) | |||
| 4.39 ± 0.8 | (62) | ||||
| umbilical artery | human | 80–120 | 5.8 ± 3.1 | (61) | |
| heterograft | bovine | 2.6 ± 0.3 | (63) |