Table 2.
The validation data for the determination of phyllanthin and hypophyllanthin by elaborated HPLC–UV method.
Validation parameters | Phyllanthin | Hypophyllanthin |
---|---|---|
Linearity (linear regression equation, regression coefficient) | y = 0.000123995x + 0.295; r = 0.9998 | y = 0.00016344x + 1.727; r = 0.9996 |
Limit of detection(LOD) [ng/ml] | 105.9 | 138.6 |
limit of quantification (LOQ) [ng/ml] | 353 | 462 |
Intra-day precission [RSD, %] | 0.38 | 0.17 |
Inter-day [RSD, %] | 0.46 | 0.48 |
repeatability [RSD, %] | 2.98 | 2.95 |