Skip to main content
. 2022 Jun 28;11(13):3723. doi: 10.3390/jcm11133723

Table 2.

Deviations (mean ± standard deviation (SD) [µm]) of the pooled data of linear distances (D1_2, D1_3, D1_4, D2_3, D2_4, D3_4) of six impression techniques (CS = CS 3800, IT = iTero Element 5D, ME = Medit i700, PS = Primescan, TR = Trios 4, CVI = conventional impression) for all groups and statistical analysis for trueness (upper right part) and precision (lower left part, presented in bold type) according to ISO 5725 [6].

Impression Technique Group Mean (Trueness) ± SD (Precision) [µm] Group A Group B Group C
CS A 87.1 ± 51.6 - 0.237 0.533
B 107.3 ± 71.1 0.374 - >0.999
C 102.2 ± 60.8 0.108 0.639 -
IT A 101.5 ± 97.1 - >0.999 0.784
B 90.5 ± 78.3 0.118 - 0.502
C 75.0 ± 96.4 0.986 0.163 -
ME A 61.5 ± 81.7 - 0.881 >0.999
B 62.2 ± 66.0 0.695 - 0.314
C 49.9 ± 48.4 0.187 0.295 -
PS A 60.7 ± 55.1 - 0.649 >0.999
B 53.6 ± 49.6 0.302 - >0.999
C 52.2 ± 40.4 0.475 0.723 -
TR A 69.4 ± 79.3 - >0.999 0.814
B 55.0 ± 53.4 0.013 - >0.999
C 50.6 ± 64.3 0.746 0.041 -
CVI A 30.5 ± 31.2 - 0.012 >0.999
B 43.5 ± 30.4 0.179 - 0.020
C 30.3 ± 25.3 0.397 0.009 -