Skip to main content
. 2022 Jul 8;12:11694. doi: 10.1038/s41598-022-15380-7

Table 2.

Basic calibrations of NMR and DNP.

Step Operation Duration Results and comments
1

Calibration of NMR frequency using 63Cu

Tune and match the NMR coil to estimated ν63Cu Record 63Cu NMR signal and refine ν63Cu

Calculate B0, ν13C, ν15N, νe−

2 min

ν63Cu = 75.259 MHz

63Cu signal originates from the coil itself, no sample needed

2

Fine calibration of NMR frequency ν13C or ν15N or other νX

Insert DNP sample

Tune and match NMR coil to the desired nucleus

Perform microwave frequency sweep

Use hyperpolarized NMR signal to adjust RF transmitter

ν13C = 71.492 MHz for pyruvate-radical concentrate

ν15N = 28.8225 MHz for urea-radical concentrate

3

Optimize µW frequency νµW and power pwMW

Insert DNP sample (or leave it after step 2 inside)

Tune and match NMR coil to desired nucleus

Perform µW frequency and power sweep

ca. 60 min
4

Optimize sample position xs inside VTI

Insert DNP sample (or leave it after step 2 inside)

Acquire NMR signal for different sample positions

ca. 30 min Optimal position xs found at 10 mm above bottom; closer to bottom assures colder sample temperature
5

Calibration of the NMR flip angle α

Insert DNP sample and adjust coil

Perform DNP until sufficient signal is obtained

Apply train of low flip angle pulses

Fit signals to obtain α for the given pdRF,paRF

Several iterations may be needed to fulfill low flip angle condition. Depending on the sample position in the coil and VTI (Step 4)

Note that the results of some steps are dependable on another and may need to be repeated (e.g. No. 3–5).