. 2022 Jul 8;12:11694. doi: 10.1038/s41598-022-15380-7

Table 5.

Standard operational procedure for dDNP experiments.

No Operation Duration Comments Norm values
1 Check T of VTI (P1, P2)

T = 1.38–1.4 K

P1 ≈ 1.3 mBar

P2 ≈ 250–350 mBar

2 Warm up stock CA-radical concentrate until liquid 10 min Protect from light
3 Flush fluid path with air 10 min Automatic
4 Add CA-radical concentrate to cup 5 min Use micropipet, add to the bottom, avoid smudging to walls

18–19 µL ≈ 22 mg (for 13C)

And 50 uL ≈ 50 mg (for 15N)

5 Prepare liquid nitrogen (lN2) 2 min Place non-magnetic dewar next to the magnet < 500 mL of lN2
6

Connect sample cup to polarizer:

- Connect empty cup

- Flush with He

- Disconnect cup, connect cup with a sample

- Dip cup in lN2

- Pressurize fluid path

3 min

Mark vial to avoid damage to thread by over tightening

Use cryogenic gloves

Immerse slowly into lN2

Pressurize while in lN2

Immersion into lN2 for 30 s
20 s
7 Insert cup into VTI 5 min

Hurry but stay calm:

open airlock, remove plug, insert sample, close airlock, start automatic insertion procedure

Manually guide tube if needed. Use gloves

8 Position cup in VTI/NMR coil Mark lowest position on tube e.g. with tape 10 mm above lowest position
9 Check that the NMR coil is in resonance 2 min Adjust tuning and matching capacitors Tune within ± 20 kHz, match 1–2%
10 Wait for temperature T and pressure to stabilize (P1, P2) 5 min Insertion causes temperature and pressure increase Close to starting temperature, e.g. T = 1.38 – 1–4 K
11

Start DNP

11a: start monitoring 13C NMR (SpinIt)

11b: start DNP (SpinAligner)

60 min

Check that µW is on

Check 13C NMR signal

12 Prepare MRI/NMR Adjust resonance of the probe, create a new protocol, shims of the system, create new project, free space to ease transport
13 Inject dissolution medium in heater Standard dissolution medium suitable for tracer 5 mL
14 Initiate heating of dissolution medium 5 min Close the cover of polarizer Automatic, ready when 11 bar reached
15 Prepare receiver vessel/syringe Weigh and label NMR tubes and place receiver in the stray field
16 Put on safety equipment: Goggles, gloves Caution! Hot liquids, high pressures!
17 Raise cup 30 s Raise cup above lHe level to reduce lHe boil off Rise by 8 cm
18 Execute dissolution < 30 s Hot liquid at high pressure and speed will be ejected Automatic
19 Collect hyperpolarized medium e.g. to syringe and apply 5 s Draw in syringe for manual injection, transfer via tubes to detection site
20 Transfer NMR samples or syringe to NMR/MRI and execute pulse sequences when sample is at the detection position 30 s A fast transfer is necessary
21 Take out the sample cup, conserve the airlock, perform the cleaning 15 min Wash the heater 3 times with deionized water, then dry at least 10 min 10 mL of water to clean the system each time