. 2022 Jun 27;10:918617. doi: 10.3389/fbioe.2022.918617

TABLE 2.

Kinematic metrics, correlations, intraclass correlations, and Bland-Altman analysis indicating bias and limits of agreement between the T’ena sensor and Vicon motion capture system.

Vicon Mean (SD) T’ena Mean (SD) ICC(2,1) Pearson’s r Mean difference Lower limit of agreement Upper limit of agreement
Block
 Movement Time 2,598.10 (327.47) 2,638.05 (333.53) 0.836 (0.808–0.860)* 0.842* −40.49 −403.90 322.88
 Path Length 1,174.89 (143.43) 1,197.60 (155.29) 0.932 (0.883–0.956)* 0.947* −23.58 −122.09 74.92
 Lift Phase Peak Velocity 11.19 (1.45) 11.10 (1.45) 0.887 (0.871–0.900)* 0.888* 0.09 −1.25 1.44
 Lower Phase Peak Velocity 10.84 (1.27) 11.53 (1.53) 0.773 (0.769–0.901)* 0.882* −0.69 −2.11 0.73
Drink
 Movement Time 3,683.15 (368.39) 3,761.05 (340.77) 0.809 (0.729–0.860)* 0.846* −77.89 −492.89 337.10
 Path Length 1,068.43 (139.47) 1,117.74 (165.69) 0.675 (0.555–0.756)* 0.721* −49.31 −278.00 179.39
 Lift Phase Peak Velocity 7.30 (1.70) 6.96 (1.16) 0.852 (0.710–0.912)* 0.891* 0.34 −0.83 1.51
 Lower Phase Peak Velocity 7.31 (1.70) 6.50 (1.23) 0.702 (0.259–0.853)* 0.845* 0.80 −1.02 2.62
Finger to Nose
 Movement Time 2,604.07 (349.42) 2,543.15 (333.80) 0.719 (0.672–0.758)* 0.731* 60.29 −431.17 553.00
 Path Length 1,058.28 (103.31) 1,105.43 (208.74) 0.496 (0.420–0.561)* 0.648* −47.15 −364.86 270.57
 Lift Phase Peak Velocity 11.56 (1.85) 12.77 (2.01) 0.717 (-0.007-0.895)* 0.894* −1.21 −3.01 0.59
 Lower Phase Peak Velocity 8.90 (1.58) 9.46 (2.11) 0.530 (0.209–0.687)* 0.680* −0.56 −3.74 2.63

* refers to statistical significance < 0.05.