Table 1.
Patient | Diaphragm motion (mm) | 4DMRI libraries | AR accuracy (mm) b | LSTM accuracy (mm) b | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Inhale phases | Exhale phases | Prediction | Matched | Prediction | Matched | ||||||||
AVG a | STD | t (s) | AVG a | STD | AVG a | STD | t (s) | AVG a | STD | ||||
1 | 15.8 | 89 | 122 | 0.24 | 0.20 | 0.44 | 0.30 | 0.20 | 0.42 | 0.28 | 1.47 | 0.47 | 0.32 |
2 | 12.9 | 97 | 118 | 0.26 | 0.20 | 0.44 | 0.31 | 0.17 | 0.59 | 0.55 | 1.46 | 0.62 | 0.58 |
3 | 20.7 | 108 | 110 | 0.44 | 0.36 | 0.44 | 0.47 | 0.36 | 0.88 | 0.70 | 1.50 | 0.93 | 0.70 |
4 | 14.4 | 108 | 111 | 0.45 | 0.32 | 0.44 | 0.48 | 0.35 | 0.61 | 0.39 | 1.49 | 0.64 | 0.38 |
5 | 9.6 | 101 | 117 | 0.25 | 0.19 | 0.44 | 0.31 | 0.18 | 0.24 | 0.21 | 1.45 | 0.27 | 0.21 |
6 | 15.8 | 114 | 102 | 0.33 | 0.28 | 0.44 | 0.47 | 0.49 | 0.64 | 0.61 | 1.47 | 0.65 | 0.63 |
7 | 11.6 | 95 | 122 | 0.42 | 0.24 | 0.44 | 0.42 | 0.26 | 0.75 | 0.48 | 1.45 | 0.78 | 0.51 |
8 | 29.1 | 92 | 125 | 0.47 | 0.49 | 0.44 | 0.63 | 0.49 | 0.81 | 1.34 | 1.45 | 0.97 | 1.24 |
AVG | 16.2 | 100.5 | 115.9 | 0.36 | 0.29 | 0.44 | 0.42 | 0.31 | 0.62 | 0.57 | 1.47 | 0.67 | 0.57 |
STD | 6.2 | 8.8 | 7.7 | 0.10 | 0.10 | 0.00 | 0.11 | 0.13 | 0.21 | 0.35 | 0.02 | 0.23 | 0.32 |
The AR (autoregression) and LSTM (long short-term memory) prediction accuracy and performance, diaphragm motion, two TR-4DMRI libraries in the inhalation and exhalation, and the library matching errors are provided. The AR method provides more accurate prediction results using less time than that of the LSTM method.
ap-value <0.02 for both prediction and matched accuracy between AR and LSTM.
bNo significant difference between prediction and matched accuracy using either AR or LSTM.