. 2022 May 25;29(8):1372–1380. doi: 10.1093/jamia/ocac072

Table 3.

Manufacturer recommended LOINC® codes that do not match LOINC® codes used in medical center pilot sitesa

Site IVD manufacturer and device name Pilot site data
Manufacturer data for similar test
Potential reason for mismatch
LIS test name LOINC® Pilot test description LOINC® LOINC® long common name
Pilot Site 1
  • Roche

  • Cobas

Calcium, urine 18488-7 Calcium 17862-4 Calcium [Mass/volume] in Urine LOINC® is for 24 h urine –the IVD manufacturer/instrument may not be aware
Cannabinoid screen 70143-3 Cannabinoids 8172-9 Cannabinoids [Presence] in Serum or Plasma by Screen method Quantitative LOINC® vs qualitative by the manufacturer
Prolactin, serum 2842-3 Prolactin 20568-2 Prolactin [Mass/volume] in Serum or Plasma by Immunoassay Methodless LOINC®—still a valid LOINC®, but not as granular as the manufacturer LOINC®
Pilot Site 2 Abbott Architect i2000 Hepatitis C Ab 16128-1 Hepatitis C virus Ab [Presence] in Serum 13955-0 Hepatitis C virus Ab [Presence] in Serum or Plasma by Immunoassay Methodless LOINC®—still a valid LOINC®, but not as granular as the manufacturer LOINC®
HBsAb, quantitative 16935-9 Hepatitis B virus surface Ab:ACnc:Pt:Ser:Qn 5193-8 Hepatitis B virus surface Ab [Presence] in Serum by Immunoassay Quantitative LOINC® vs qualitative by the manufacturer
Pilot Site 3
  • Roche

  • Cobas c 702

Blood urea nitrogen 3094-0 BUN 14937-7 Urea nitrogen [Moles/volume] in Serum or Plasma Different units of measure: mass/volume versus moles/volume
AST (SGOT) 1920-8 Aspartate aminotransferase (AST) 30239-8 Aspartate aminotransferase [Enzymatic activity/volume] in Serum or Plasma by With P-5'-P Methodless LOINC®—still a valid LOINC®, but not as granular as the manufacturer LOINC®
Pilot Site 5
  • Roche

  • Cobas c 702

BUN 3094-0 Urea nitrogen [Mass/volume] in Serum or Plasma 14937-7 Urea nitrogen [Moles/volume] in Serum or Plasma Different units of measure: mass/volume versus moles/volume
Aspartate aminotransferase (AST) 1920-8 Aspartate aminotransferase [Enzymatic activity/volume] in Serum or Plasma 30239-8 Aspartate aminotransferase [Enzymatic activity/volume] in Serum or Plasma by With P-5'-P Methodless LOINC®—still a valid LOINC®, but not as granular as the manufacturer LOINC®
a

We identified 136 LOINC® codes in the manufacturer LIVD files that mismatch LOINC® codes used at pilot site medical centers. This table shows a representative sample of these mismatches.