Table 2. Validation Parameters of the Developed TLC Visualization Method to Determine LOP and BIS in their Pure Form.
parameters | LOP | BIS |
---|---|---|
concentration range (μg/mL) | 2.00–10.00 | 1.00–10.00 |
linearity (regression equation) | Y = −2.815x + 52.85 | Y = −3.6595x + 47.673 |
correlation coefficienta | 0.9999 | 0.9996 |
accuracy (mean ± SD)b | 100.04 ± 0.48 | 99.93 ± 1.46 |
LOD (μg/mL)c | 0.57 | 0.10 |
LOQ (μg/mL)c | 1.73 | 0.30 |
Average of three determinations.
Average of 5 different concentrations of each LOP and BIS.
LOD is calculated via (SD of response/slope) × 3.3 and LOQ is via (SD of response/slope) × 10.
Interday precision; the RSD of 3 different concentrations; 4.00, 6.00, and 8.00 μg/mL for LOP and 2.00, 4.00, and 6.00 μg/mL for BIS; 3 replicates each; on the same day.
Interday precision; the RSD of 3 different concentrations; 4.00, 6.00, and 8.00 μg/mL for LOP and 2.00, 4.00, and 6.00 μg/mL for BIS; 3 replicates each; on 3 successive days.