Skip to main content
. 2022 Jun;63(3):224–230. doi: 10.3325/cmj.2022.63.224

Table 4.

Quality metrics for sequencing runs compared to assess library normalization methods. In runs 1-3 libraries were normalized with magnetic beads included in the Nextera® XT library prep kit, while in runs 4-6 individual dilution was applied after quantification with LabChip® DNA High Sensitivity Assay on LabChip® GX Touch HT

Run No. of samples Normalization method Cluster density (K/mm2) Clusters PF (%) % bases≥Q30 % reads identified ± standard deviation
(expected %) % CV
1
24
Bead-based
939
94.1
93.0
4.0 ± 1.3 (4.2)
32.1
2
24
491
97.8
97.1
3.9 ± 1.8 (4.2)
47.1
3
24
1062
91.5
93.8
4.1 ± 1.6 (4.2)
40.0
4
24
LabChip-based 864
93.6
93.4
4.0 ± 0.6 (4.2)
16.1
5
28
948
93.7
96.1
3.7 ± 0.7 (3.6)
19.4
6 24 551 96.0 95.7 4.1 ± 1.7 (4.2) 41.7