Skip to main content
. 2022 May 23;2022:244–253.

Table 2.

VitalSeer 1.0 and 1.2 clinical studies – HR data analysis summary.

Studies Readings with confidence level >90% Mean Error (bpm) RMSE (bpm) MAE (bpm) % of values within 2 bpm % of values within 5 bpm Outlier ratio (error >10 bpm) %
Aggregated VitalSeer 1.0 and 1.2 studies with integrated webcam 86.1% (n=15285 predictions) -0.07 1.20 0.74 93.1 99.3 0.05
Pilot with FLIR Blackfly camera (concurrent to VitalSeer 1.2 study) 97.2% (n=9843) -0.06 1.09 0.68 94.6 99.3 0.01