Skip to main content
. 2021 Oct 4;56(11):e4786. doi: 10.1002/jms.4786

TABLE 1.

Measured size data, theoretical MW data, derived MW data and mass spectrometric data of empty and filled AAV8 VLP preparation

AAV8 VLP preparations
Investigative approach: Empty Filled
nES GEMMA EMD data (nm) 25.10 ± 0.18 25.93 ± 0.07
Theoretical MW based on 1:1:10 VPs ratio (kDa) a 3746 5076
Theoretical MW based on SDS‐PAGE (kDa) 3658 4988
EMD/MW correlations (kDa) 3670 ± 69 4751 ± 47
Native MS (kDa) (for n = 161 charges) 3710 5005

Abbreviations: EMD, electrophoretic mobility diameter; nES GEMMA, native nano‐electrospray gas‐phase electrophoretic mobility molecular analyzer; MS, mass spectrometry.

a

Based on the following MW: VP1 81 kDa; VP2 65 kDa; VP3 60 kDa. 48