. 2022 Jul 14;11(7):12. doi: 10.1167/tvst.11.7.12

Table 3.

Agreement Calculation of Measured Vascular Features Between AutoMorph and Expert Annotation

ICC (95% Confidence Interval)
Zone B Zone C Whole Image
DR HAGIS
 Fractal dimension 0.94 (0.88–0.97) 0.98 (0.95–0.99) 0.94 (0.88–0.97)
 Vessel density 0.98 (0.96–0.99) 0.97 (0.94–0.99) 0.94 (0.88–0.97)
 Average width 0.95 (0.89–0.98) 0.96 (0.93–0.98) 0.97 (0.95–0.99)
 Distance tortuosity 0.80 (0.59–0.91) 0.85 (0.69–0.93) 0.86 (0.73–0.93)
 Squared curvature tortuosity 0.68 (0.34–0.85) 0.88 (0.75–0.94) 0.84 (0.68–0.92)
 Tortuosity density 0.89 (0.77–0.95) 0.70 (0.38–0.86) 0.87 (0.74–0.93)
IOSTAR-AV
 CRAE (Hubbard) 0.81 (0.56–0.92) 0.82 (0.57–0.91)
 CRVE (Hubbard) 0.8 (0.54–0.91) 0.78 (0.52–0.89)
 AVR (Hubbard) 0.87 (0.69–0.94) 0.81 (0.66–0.92)
 CRAE (Knudtson) 0.76 (0.45–0.9) 0.75 (0.44–0.89)
 CRVE (Knudtson) 0.85 (0.67–0.94) 0.86 (0.58–0.9)
 AVR (Knudtson) 0.85 (0.66–0.94) 0.82 (0.51–0.91)

The agreement of vessel caliber was validated on the IOSTAR-AV dataset, other metrics with the DR HAGIS dataset. Because caliber features rely on the six largest arteries and veins in Zones B and C, there is no caliber feature for the whole image level.