Performance of antigen detection tests compared with viral culture

Culture POS Culture NEG Statistic Value 95% CI
LumiraDx antigen results versus viral culture
Lumira POS 101 23 Sensitivity 0.90 0.83–0.94
Lumira NEG 11 54 Specificity 0.70 0.59–0.79
BD Veritor antigen results versus viral culture
BD POS 82 6 Sensitivity 0.74 0.65–0.81
BD NEG 29 70 Specificity 0.92 0.84–0.96
CareStart antigen results versus viral culture
CareStart POS 80 7 Sensitivity 0.74 0.65–0.81
CareStart NEG 28 69 Specificity 0.91 0.82–0.95
Oscar Corona antigen results versus viral culture
Oscar POS 75 6 Sensitivity 0.74 0.65–0.82
Oscar NEG 26 69 Specificity 0.92 0.84–0.96