. 2022 Jan 19;16(3):788–789. doi: 10.1177/19322968211070815

Table 1.

Accuracy Results for Bolus and Basal Rate Delivery.

Bolus delivery (n = 225) Basal delivery (648 h)
0.1 U 0.2 U 1 U 0.1 U/h 1 U/h
Mean deviation 1.3% −2.5% −0.3% +7.9% +1.2%
SD of deviation for boluses/1-h windows 12.9% 18.5% 13.6% 21.7% 10.0%
Boluses/1-h windows within ±15% 76.4% 61.3% 71.6% 52.7% 86.2%
Accuracy limits of manufacturer a ±50% ±25% ±5% ±5% ±5%
Boluses/1-h windows within accuracy limits of manufacturer 100.0% 80.4% 40.4% 19.2% 40.7%

The mean and standard deviation (SD) summarize the relative deviation from the target delivery of individual boluses (bolus delivery) and nonoverlapping 1-hour windows (basal rate delivery) over all tested disposables.

a

The accuracy limits given by the manufacturer for bolus delivery are ±5% for bolus volumes ≥1 U and ±0.05 U for volumes <1 U. The accuracy limits for basal rate delivery are ±5% for basal rates ≥0.05 U/h. 6