Table 3.

Accuracy of the In Vivo Interstitial Fluid Ketone Sensor Results Compared to the Capillary Ketone Strip Reference. Table Reproduced from Alva et al. 1

Concentration range Percentage within Mean MAD/MARD SD/%CV Number of paired data points
0.225 mM/20% 0.3 mM/30% Between subject Within subject
<1.5 mM 83.4% 91.7% 0.129 mM 0.051 mM 0.029 mM 2724
≥1.5 mM 76.0% 89.7% 14.4% 5.40% 7.46% 408
Combined 82.4% 91.4% 3132

Accuracy and bias results for ketone concentrations <1.5 mM are in mM. SD is the standard deviation and CV is the coefficient of variation.