Skip to main content
. 2021 Jan 5;16(3):677–682. doi: 10.1177/1932296820982584

Table 3.

Rebound Hyperglycemia Events in Users of the G5 and G6 Systems.

Preceding glucose nadir, mg/dL G5 G6 Relative difference (%) P value
Frequency (per week) <70 1.83 (1.30) 1.70 (1.36) 7 <.001
<55 0.78 (0.68) 0.52 (0.59) 33 <.001
Duration (min) <70 214 (142) 197 (151) 8 <.001
<55 219 (161) 171 (190) 22 <.001
AUC >180 mg/dL (mg/dL × min) <70 15 601 (17 936) 13 638 (17 527) 13 <.001
<55 16 741 (21 338) 12 283 (21 669) 27 <.001

Note. RH events were defined as series of one or more sensor glucose values >180 mg/dL starting within two hours of a sensor glucose value <70 mg/dL or a subset of these events starting within two hours of a sensor glucose value <55 mg/dL. Results given as mean (SD).

Abbreviations: AUC, area under the (glucose concentration-time) curve.