Scatter fraction as calculated by the time‐of‐flight single scatter simulation algorithm using positron emission tomography attenuation maps (μ‐maps) from computed tomography (CT), lutetium oxyorthosilicate (LSO) transmission events (Tx) before injection (b.i.), and LSO‐Tx after injection (a.i.)

CT μ‐maps (%) LSO‐Tx b.i. (%) LSO‐Tx a.i. (%)
Patient 1 38.3 37.7 38.4
Patient 2 40.3 35.1 35.1
Patient 3 40.1 40.0 38.0