TABLE 2.

Flow‐rate assessment for steady flow in a straight tube (experiment 1)

Acquisition Volumetric flow rate (mL/s) MRI‐assessed flow rate (mL/s) Relative difference (%) (ΔVolumetric) MRI‐assessed max velocity (cm/s) Relative difference (%) (Δ4DGRE) MRI‐assessed flow rate (mL/s) Relative difference (%) (ΔVolumetric) MRI‐assessed max velocity (cm/s) Relative difference (%) (Δ4DGRE)
Tube 1 (diameter 1.0 cm) Tube 2 (diameter 0.7 cm)
4DEPI5_ro_coronal 50 53.6 7.2 96 9.1 44.0 −12.0 216 −1.4
4DEPI5_bp_coronal 50 53.0 6.1 97 10.2 45.0 −10.0 216 −1.4
4DEPI5_nbp_axial 50 52.5 5.0 98 11.4 48.0 −4.0 215 −1.8
4DEPI3_ro_coronal 50 54.0 8.0 88 0.0 49.1 −1.8 233 6.4
4DEPI3_bp_coronal 50 52.3 4.6 93 5.7 50.0 0 215 −1.8
4DEPI3_nbp_axial 50 51.6 3.1 92 4.5 46.7 −6.6 229 4.6
4DGRE_coronal 50 49.4 −1.2 88 47.8 −4.4 219
4DEPI5_ro_coronal 23 24.3 5.7 44 2.3
4DEPI5_bp_coronal 23 24.4 6.1 46 7.0
4DEPI5_nbp_axial 23 23.9 3.9 42 −2.3
4DGRE_coronal 23 24.4 3.0 43

4DEPIx_ro refers to acquisition with EPI factor × (5 or 3) with the readout gradient aligned parallel to the main flow direction; 4DEPIx_bp refers to EPI acquisition with the blip phase‐encoding gradient aligned parallel to the main flow direction; and 4DEPIx_nbp refers to EPI acquisition with the readout and blip phase‐encoding gradient both perpendicular to the main flow direction.