Table 4.

Matrix effect, recovery and process efficiency. (n=8 for thiotepa and tepa, n=3 for CP and 4OHCP)

Analytes Conc. (ng/ml) Matrix effect Recovery PE
analyte IS IS-normalized analyte IS analyte IS

Thiotepa Low (3) 108 105 103 86.3 95.9 93.3 101
Med (80) 105 110 95.5 90.3 95.0 94.6 105
High (1600) 106 111 95.5 93.4 98.1 99.2 109

tepa Low (3) 121 119 102 88.3 94.5 107 112
Med (80) 120 120 100 86.3 92.7 104 112
High (1600) 111 115 96.5 89.0 93.4 98.8 107

CP Low (60) 105 98.9 106 107 103 112 102
Med (800) 101 100 101 103 101 103 101
High (8000) 108 102 106 90.2 95.9 97.6 97.5

4OHCP Low (60) 105 98 107 113 107 118 104
Med (800) 103 100 103 115 102 118 102
High (4000) 112 104 108 99.3 95.8 111 99.8