Table 1.
Linearity, accuracy, and precision data for NZ and NZ-NO by the proposed method.
Linearity results | |||||
---|---|---|---|---|---|
| |||||
Parameter | Results | ||||
|
|||||
NZ | NZ-NO | ||||
Concentration range (μg/mL) | 0.1–2.0 | 0.5–15 | |||
Limit of detection (LOD) (μg/mL) | 0.02 | 0.1 | |||
Limit of quantification (LOQ) (μg/mL) | 0.1 | 0.5 | |||
Correlation coefficient (r) | 0.9998 | 0.9999 | |||
Slope | 21.32 × 103 | 5.41 × 103 | |||
Intercept | 3.38 × 103 | 1.63 × 103 | |||
Standard deviation of the residuals (Sy/x) | 0.29 × 103 | 0.16 × 103 | |||
Standard deviation of the intercept (Sa) | 0.14 × 103 | 0.05 × 103 | |||
Standard deviation of the slope (Sb) | 0.17 × 103 | 0.01 × 103 | |||
% RSD | 1.54 | 1.37 | |||
% Error (% RSD/√n) | 0.61 | 0.52 | |||
| |||||
Accuracy and precision results | |||||
| |||||
Analyte | Conc. (μg/mL) | Intra-day (n = 5) | Inter-day (n = 5) | ||
|
|
||||
Accuracy (%) | Precision (% RSD) | Accuracy (%) | Precision (%RSD) | ||
| |||||
NZ | 0.1 | −1.06 | 2.36 | −3.61 | 3.91 |
0.6 | −2.35 | 2.13 | −3.04 | 2.50 | |
1.5 | +0.11 | 1.84 | +1.65 | 2.56 | |
NZ-NO | 0.5 | +0.68 | 2.93 | +1.39 | 3.90 |
3.5 | +3.38 | 2.00 | +2.01 | 3.96 | |
15 | +1.52 | 1.45 | +3.71 | 2.20 |