Skip to main content
. 2019 Jun 26;27(4):906–914. doi: 10.1016/j.jfda.2019.06.001

Table 2.

Linearity, precision and accuracy for 12 compounds.

Compound Calibration curve range (ng/mL) R2 Precision (RSD %) Accuracy (Error %)


Intraday Interday Intraday Interday
Amlodipine 15–7500 0.999 2.5–8.1 6.2–9.5 4.7–28.1 4.1–23.1
Atenolol 25–12500 0.999 4.3–5.9 7.8–9.1 6.0–13.2 1.2–17.8
Atorvastatin 1–500 0.991 2.2–6.3 3.7–13.1 8.7–18.3 5.1–15.3
Digoxin 10–5000 0.996 2.2–5.9 5.9–8.8 5.6–13.4 7.5–8.6
Enalapril 0.5–250 0.993 1.3–6.7 5.8–8.5 1.0–6.9 5.2–6.2
Fenofibrate 15–7500 0.992 1.1–9.7 10.2–13.0 8.5–15.7 5.7–12.1
Furosemide 100–50000 0.998 2.2–5.2 4.2–7.9 0.7–13.1 4.9–12.1
Losartan 1–500 0.996 2.0–5.0 4.5–9.2 7.3–17.1 4.6–12.4
Nifedipine 25–12500 0.998 2.7–4.7 6.8–7.7 1.2–9.5 6.3–7.6
Propranolol 1–500 0.996 5.4–9.0 6.7–9.2 1.8–4.3 2.5–6.1
Simvastatin 25–12500 0.992 3.9–6.9 4.7–6.9 2.5–9.5 5.2–7.4
Valsartan 5–2500 0.995 3.4–4.9 4.7–8.9 3.7–5.4 5.2–6.0