表 2.
比较使用Autocal校准MLC叶片到位精度前后3种验证方式结果
Comparison of 3 dose verification methods before and after MLC calibration using Autocal software (Mean±SD)
| Autocal time | Mapcheck | Arccheck | 3DVH | |||||
| 3%/3 mm | 2%/2 mm | 3%/3 mm | 2%/2 mm | 3%/3 mm | 2%/2 mm | |||
| P1: Mapcheck (After autocal) vs Arccheck (After autocal); P2: Arccheck (After autocal) vs 3DVH (After autocal); P3: 3DVH (After autocal) vs Mapcheck (After autocal). | ||||||||
| Before autocal | 88.80±1.81 | 85.05±1.90 | 87.61±1.98 | 83.78±2.05 | 89.87±2.28 | 86.61±2.42 | ||
| After autocal | 99.25±0.53 | 95.81±0.69 | 98.13±0.99 | 94.25±1.05 | 98.30±1.15 | 94.30±1.12 | ||
| P | 0.001 | 0.005 | 0.003 | 0.002 | 0.002 | 0.001 | ||
| P1 | 0.021 | 0.032 | - | - | - | - | ||
| P2 | - | - | 0.011 | 0.028 | - | - | ||
| P3 | - | - | - | - | 0.008 | 0.015 | ||