. 2022 Jul 20;42(7):1089–1094. [Article in Chinese] doi: 10.12122/j.issn.1673-4254.2022.07.19

表 2.

比较使用Autocal校准MLC叶片到位精度前后3种验证方式结果

Comparison of 3 dose verification methods before and after MLC calibration using Autocal software (Mean±SD)

Autocal time Mapcheck Arccheck 3DVH
3%/3 mm 2%/2 mm 3%/3 mm 2%/2 mm 3%/3 mm 2%/2 mm
P1: Mapcheck (After autocal) vs Arccheck (After autocal); P2: Arccheck (After autocal) vs 3DVH (After autocal); P3: 3DVH (After autocal) vs Mapcheck (After autocal).
Before autocal 88.80±1.81 85.05±1.90 87.61±1.98 83.78±2.05 89.87±2.28 86.61±2.42
After autocal 99.25±0.53 95.81±0.69 98.13±0.99 94.25±1.05 98.30±1.15 94.30±1.12
  P 0.001 0.005 0.003 0.002 0.002 0.001
  P1 0.021 0.032 - - - -
  P2 - - 0.011 0.028 - -
  P3 - - - - 0.008 0.015