Skip to main content
. 2022 Mar 22;37(6):1537–1543. doi: 10.1111/jocs.16426

Table 1.

Procedural data of animal experiments

No. Weight Mitral annulus diameter (mm) Mitral valve area Stent size Stent area Operation time Intervention time Blood loss volume
(kg) (cm2) (mm) (cm2) (min) (min)
1 55 26 676 27 729 92 39 353
2 64 27 784 27 900 NA NA NA
3 56 28 729 30 729 75 25 304
4 59 28 784 30 900 76 21 365
5 62 30 900 33 1089 65 35 300
6 69 29 841 33 1089 70 29 433
7 70 32 1024 35 1225 93 31 310
8 72 30 900 35 1225 90 25 394
9 60 33 1089 35 1225 77 28 365
10 62 30 900 35 1225 58 14 342
Mean ± SD 62.90 ± 5.8 29.30 ± 2.05 862.70 ± 121.24 32.00 ± 3.10 1066.60 ± 193.37 77.33 ± 11.58 27.44 ± 6.99 351.78 ± 41.54

Note: No. 2 pig died of failed valve implantation after 1 day, and its data were not applicable and excluded.