Table 1.
Features | TD-OCT | SD-OCT | SS-OCT |
---|---|---|---|
Movable Reference Mirror | Mobile Reference Mirror, Spectrometer | Swept-Source Laser | |
Wavelength | 810 nm | 840 nm | 1052 nm |
Axial resolution | 10 µm | 8 µm | 6 µm |
Lateral resolution | 20 µm | 20 µm | 20 µm |
Speed of A scan per sec | 512 | 50,000 | 100,000 |
B scan measurement time | - | 1.0 s (50×) | 1.0 s (96×) |
Length of line scan | 6 mm | Up to 9 mm | Up to 12 mm |
Artifacts | more | less | less |