. 2022 Jul 10;12(7):1682. doi: 10.3390/diagnostics12071682

Table 4.

Diagnostic performance of linearly blended M_0.6, standard virtual monoenergetic imaging (VMI) at 60 keV, and noise-optimized VMI+ series at 40 keV and 60 keV levels for the detection of portal vein thrombosis (PVT).

Parameter M_0.6 60 keV VMI 40 keV VMI+ 60 keV VMI+
Sensitivity 87%
[0.77–0.94]
87%
[0.77–0.94]
96%
[0.89–0.99]
92%
[0.84–0.97]
Specificity 92%
[0.87–0.96]
97%
[0.93–0.99]
96%
[0.93–0.99]
97%
[0.93–0.99]
AUC 0.94
[0.90–0.97]
0.95
[0.91–0.98]
0.99
[0.98–1.00]
0.97
[0.93–0.99]
p-value p < 0.001 p = 0.001 N/A p = 0.012

Sensitivity and specificity are presented as percentages with corresponding 95% confidence intervals (CI) in brackets. VMI = virtual monoenergetic imaging, VMI+ = noise-optimized virtual monoenergetic imaging, AUC = Area under the curve. p-values refer to the comparison with 40 keV VMI+ values.