Skip to main content
. 2022 Jun 27;58(7):858. doi: 10.3390/medicina58070858

Table 1.

Study parameters and outcomes. SS = Stainless Steel, Ti = Titanium, R = right, L = left, RIL = relative implant load and LSB = least significant bits. * mRUST scoring not performed.

Animal ID Side Plate
Material
Gap Width Monitoring Period Loading Events Active Time Time to 50% RIL RIL End Level Device Sensitivity (Plate Specific)
mm days total avg/day avg %/day days % LSB/kg
519030 * R SS 0.6 227 119,462 872 5.9 31 21 16.4
519042 L Ti 0.6 252 194,460 1350 6.07 30 16 31.6
519031 * L SS 2 207 147,350 1002 7.4 23 11 14.6
519043 R Ti 2 232 127,981 908 5.88 34 9 36.2
519051 R SS 4 245 181,762 1387 7.29 24 10 15.9
519049 L Ti 4 197 123,954 1158 5.32 22 11 24.7
519050 L SS 6 250 271,940 1915 6.71 63 15 15.0
519052 R Ti 6 245 146,161 981 7.37 45 3 27.2
519032 * R SS 10 209 205,236 1567 7.5 20 16 14.2
519044 L SS 10 216 255,216 1701 6.62 82 11 18.6
520001 R SS 30 255 213,563 1396 5.54 234 56 16.4
mean 230 180,644 1294 6.51 55 16 21.0
SD 21 52,730 343 0.81 62 14 7.7
min 197 119,462 872 5.32 20 3 14.2
max 255 271,940 1915 7.50 234 56 36.2