Table 3.

Computed tomography dose index (CTDIvol), dose length product (DLP), and effective dose of all patients examined with the photon counting CT (PCCT) and with the standard CT (S40). Mean ± SD.

CTDIvol
(mGy)
DLP
(mGy*cm)
Eff. Dose
(mSv)
PCCT S40 p PCCT S40 p PCCT S40 p
Total 3.8 ± 1.2 5.2 ± 1.5 <0.001 158.2 ± 58.0 232.5 ± 67.6 <0.001 2.4 ± 1.0 3.4 ± 1.0 <0.001
Total,
BMI < 30
3.4 ± 0.8 4.5 ± 1.1 <0.001 138.3 ± 33.7 203.9 ± 62.5 <0.001 2.1 ± 0.6 3.1 ± 1.0 <0.001
Male 3.5 ± 0.7 5.1 ± 1.3 <0.001 144.7 ± 37.8 231.2 ± 55.8 <0.001 2.0 ± 0.5 3.3 ± 0.8 <0.001
Female 4.1 ± 1.5 5.3 ± 1.8 0.080 170.8 ± 71.0 235.0 ± 89.0 0.030 2.8 ± 1.2 3.8 ± 1.4 0.034