Table 1.

Image acquisition protocol

TR/TE (ms) Voxel size (mm3) Acquisition matrix Reconstruction matrix Receiver bandwidth Hz/px Flip angle b-values (s/mm2) Tumour coverage
Pre-contrast
 High resolution 2-point 3D GRE DIXON 7/3.69 1.25 × 1.25×2 192 × 154 192 × 192 605 four o/16o N/A Full
 DW-EPI 11200/77 1 × 1×4 128 × 104 208 × 256 2055 90o 50, 300, 600, 900 Full
 2D multi echo GRE 100/10 echo times (range 4.9–68.9) 1.5 × 1.5×3 160 × 160 160 × 160 490 45 N/A Central volume
Post-contrast
 DCE-MRI 2-point 3D GRE DIXON 5.74/2.46 1.5 × 1.5×3 160 × 141 160 × 160 600 four o/16o N/A Central volume
 High resolution 2-point 3D GRE DIXON 7/3.69 1.25 × 1.25×2 192 × 154 192 × 192 605 four o/16o N/A Full

DCE-MRI, Dynamic contrast enhanced-MRI; DW-EPI, diffusion weighted echoplanar imaging; GRE, Gradient echo.

DWI, 2D multiecho GRE and DCE-MRI sequences were centered on a slice chosen by a specialist sarcoma radiologist to ensure targetingthe section of the tumour with increased cellularity and to avoid areas of high fat content.

DCE-MRI comprised 85 dynamic acquisitions with time resolution 3 s with the highest flip angle (16o) [a gadolinium based contrast administration (2 ml s−1 followed by 20 ml saline flush)]