TABLE 7.
Contribution of measurement uncertainty in the purity analysis of the original La2O3 sample.
Uncertainty component | Relative standard uncertainty, % |
---|---|
Repeatability of the gravimetric analysis | 0.0080 |
Stoichiometry of weighing forms | 0.024 |
Weighing for original La2O3 | 0.00066 |
Weighing for oxide form | 0.00069 |
Weighing for sulfate form | 0.00069 |
Impurities | 0.000019 |
Molecular weights | |
La2O3 | 0.00040 |
La2(SO4)3 | 0.0053 |
Combined standard uncertainty | 0.026 |
Coverage factor | 2.18 |
Expanded uncertainty(U), relative (%) | 0.057 |