Table 3.
Dose measurements for different settings.
Covered | Voltage [kV] | Current [A] | FOV [mm] | Pixel size [m] | Time [s] | Dose [mGy] |
---|---|---|---|---|---|---|
– | 90 | 200 | 73 73 | 143 | 19 | 14.0 |
– | 90 | 200 | 73 73 | 143 | 19 | 13.9 |
– | 90 | 200 | 73 73 | 143 | 19 | 14.8 |
– | 90 | 200 | 40 40 | 78 | 19 | 21.3 |
– | 90 | 200 | 20 20 | 39 | 19 | 41.6 |
x | 90 | 200 | 73 73 | 143 | 19 | 13.8 |
x | 90 | 200 | 40 40 | 78 | 19 | 20.0 |
Measurements taken with wrapping the probe with a layer of 1cm meat to mimic scattering within a mouse are indicated by ’x’. Please note that a smaller FOV (smaller pixel size) is achieved by moving the X-ray source closer to the mouse/specimen which results in elevated dose levels.